ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacture's declaration - electromagnetic emissions
The AN-6019 is intended for use in the electromagnetic environment specified below.
The customer or the user of the AN-6019 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The AN-6019 is intended for use in the electromagnetic environment specified below. The
customer or the user of the AN-6019 should assure that it is used in such an environment.
IMMUNITY
IEC 60601test
test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Compliance
Group 1
Class B
Not
applicable
Not
applicable
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Compliance
level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic environment-guidance
The AN-6019 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment
The AN-6019 is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Electromagnetic
level
environment-guidance
Floors should be wood, concrete or
ceramic tile.If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.