SET-UP & OPERATION
1
2
Fully charge battery for 10-17
Insert container into holder.
hOURS.
NOTE–Inspect suction tubing and container for leaks, cracks, etc. and assure that all connections are secure and without leaks before using.
7
8
If using AC or DC power, plug
Plug the other end into an AC
the small connector into the
wall outlet or DC receptacle.
DC power input on the side of
NOTE– The AC adapter may
unit.
become warm to the touch
during charging or running of
the unit. This is normal.
wARNING
If the unit is not receiving power from an external source or the battery was not recharged, the low battery indicator light will remain on and the performance of
the unit will drop off rapidly. Switch to another power source immediately after the low battery light appears to avoid an interrupted suction procedure.
NOTE– Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be checked.
CAUTION– When automatic float shut-off is activated, contents of the collection container should be emptied. Further suctioning could cause damage to the vacuum pump.
CAUTION– Should fluid be aspirated back into the unit, equipment provider servicing is necessary as possible vacuum pump damage may result.
bATTERY CHARGING & FILTER MAINTENANCE
battery Charging
1
2
Ensure power switch is
"off". Plug the small connector of
the AC or DC adapter into the
DC power input.
NOTE– A discharged battery will require 10-17 hours (depending on depth of discharge) of charging to reach a full capacity.
NOTE– Do not connect the AC adapter to an outlet controlled by a switch to ensure power is supplied to unit at all times.
NOTE– Do not connect the DC power cord to an outlet that is not constantly energized.
NOTE– A fully charged battery will provide approximately 60 minutes of continuous operation at a zero vacuum level (free flow). Operation time will decrease with higher
vacuum levels.
NOTE– If unit is not in use for extended periods, battery should be recharged every 3 months minimum.
CAUTION– Discharging the battery completely will shorten the life of the battery. Do not operate the unit more than a few minutes if the low battery indicator light is lit.
Recharge as soon as possible.
NOTE– When charging the battery, use an external power source and verify that the charge light illuminates when the unit is in the "Off' position. If the battery does not
charge, please contact your authorized DeVilbiss Healthcare provider.
NOTE– The internal rechargeable battery is sealed lead-acid. Contact your local authorities for instruction on proper disposal.
EN - 4
3
4
Attach 43/8" tubing from filter
Attach 6' patient tubing to
cartridge to tubing connector.
container lid at outlet labeled
<Patient>.
9
10
Adjust the suction level.
Turn the unit
"on".
3
4
Plug the other end into an AC
Battery charging begins; 10-17
wall outlet or DC receptacle.
hours for full charge.
5
6
Select desired power source.
Ensure power switch is
"off".
(Skip steps 7-8 if using internal
battery power.)
11
Verify suction level.
NOTE - Always verify suction
level before beginning by
occluding open end of patient
tubing while observing gauge.
Adjust knob to desired level.
5
Battery charging complete.
A-7314