TROUbLESHOOTING
NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO YOUR DEVIlBISS
HEAlTHCARE PROVIDER WHERE YOU PURCHASED IT, PlEASE TAkE A FEW mOmENTS TO CHECk FOR THESE POSSIBlE CAUSES:
PRObLEM
Unit does not turn on (no external power is connected).
Unit does not turn on when external power is connected.
Green external power light does not illuminate.
Pump runs, but there is no suction.
Low suction.
Battery will not charge (external power and charge
indicator lights should be illuminated during charge
mode)
SPECIFICATIONS/CLASSIFICATIONS
Size (including container)
Weight (including container)
Typical Operating Sound Level
Electrical Requirements
Vacuum Range
Air Flow @ Pump Inlet:
Collection Container Capacity
Warranty
Internal Battery
Approvals
Adapter Manufacturer Information
Environmental Conditions
Operating Temperature Range
Operating Relative Humidity
Operating Atmospheric Pressure
Storage & Transport Temperature Range
Storage & Transport Relative Humidity
Storage & Transport Atmospheric Pressure
Equipment Classifications
With respect to protection from electric shock
Degree of protection against electric shock
Degree of protection against ingress of liquids
Mode of Operation
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
Electrically powered medical suction equipment for field and transport use according to EN ISO 10079-1 : 2009
High Flow/High Vacuum
* Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
Manufacturer's Note
Thank you for choosing a DeVilbiss Suction Unit. We want you to be a satisfied customer. If you have any questions or comments, please send them to our address on the
back cover.
For Service Call Your Authorized DeVilbiss Healthcare Provider:
Phone
DEVILbISS GUIDANCE AND MANUFACTURER'S DECLARATION
wARNING
Medical Electrical Equipment needs special precautions regarding Electromagnetic Compatibility (EMC) and needs to be installed and put into service according
to the EMC information provided in the accompanying documents.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment or
system should be observed to verify normal operation in the configuration in which it will be used.
EN - 6
ACTION
1. Check that battery is fully charged and/or charge battery.
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.
8.3 H x 8.0 W x 8.5 D (21.1 cm x 20.3 cm x 21.6 cm) (not including AC to DC adapter)
6.6 lb. (3 kg) (not including AC to DC adapter)
55 dBA
100-240V~, 50/60Hz, 1.2A max
50-550 mm Hg +/- 10%*
27 LPM (free flow) typical (may be less when running from internal battery)*
800 ml (cc) Disposable
Two-years limited, excluding internal battery and collection container
90-day
IEC 60101-1-2; CSA-C22.2 # 601.1; UL 60601-1; EN 60601-1-2; ISO 10079-1; IEC 60601-1; IEC 60529
IP12; IEC 60601-1-6; CENELEC EN 60601-1
Emerson Model # AD5012N2LM or Autec Power Systems Model # DTM36-12 or SL Power/Ault Model #
MENB1040A1240N02
32°F (0°c) - 104°F (40°c)
0-95%
10.2 psi (70 kPa) - 15.4 psi (106 kPa)
-40°F (-40°c) - 158°F (70°c)
0-95%
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Class II and internally powered
Type BF Applied Parts
IP12 and ordinary power supply
Intermittent Operation: 30 minutes on, 30 minutes off
Purchase Date
; 12V
; 33 W Max
Serial #
A-7314