icons to answer the questions.
After entering the message box, the screen displays the most
recent message. You can click Previous or Next button to
move to the previous or next message.
You can upload your answer onto the database and
download new messages from the website. Press Home
Button to return to the Main Page.
SERVICE BUTTON
The meter is equipped with a Service Button.
Pressing and holding the button for more than 3 seconds will
send a Request Assistance to the care service provider.
For more information about the Service Button, please
contact customer service or your healthcare professional.
Warning:
The Service Button is not intended for an emergency. Please
DO NOT expect an instant response or immediate assistance
by pressing this button. Your care service provider will define
the service agreement with you.
Maintenance
Battery
Your meter comes with one 3.7V Li-ion rechargeable battery.
Low Battery Signal
The meter will display the message «Low battery» with a
«Warning Symbol» to alert you when the meter power is
getting low.
The power is not enough to do a test. Please recharge the
battery immediately.
Recharging the battery
The power adapter can be used as the battery charger when
you need to recharge the Li-ion battery. To recharge the 3.7V
Li-ion battery, connect the AC adapter to the meter and a
power source.
The power indicator will light up in red, which indicates the
battery is undergoing a recharge. It should take approximately
5 hours to be fully charged. If the power indicator is blinking
in red when connecting the AC adapter to a power source,
this indicates the meter is in a pre-charging stage. If the power
indicator starts blinking in green, this indicates the recharge
is nearly finished. The power indicator will diminish when the
recharge completes.
CAUTION:
The Li-ion battery must be recharged with the FORA 6 GTel
meter by connecting the AC adapter to an electrical outlet.
Only this 3.7V rechargeable Li-ion battery (battery model no.
ICP553136ARM) can be recharged by FORA 6 GTel with the
connection of the AC adapter. Any other kind of rechargeable
battery is not allowed or it may damage the meter.
EN-16
CAUTION