Technical Information
Interconnections
Pin 3-8 must be shorted to allow the Turbo-V 2300 Rack Controller pump to start
NOTE
if no interlock contact is connected. If after starting the pump, the interlock
contact opens, it has no effect on the operation and the pump continues to turn.
The following figure shows a typical contact logic input connection
and the related simplified circuit of the controller.
Figure 10

Connection J2 - Logic Output Interconnections

Figure 11
214/262
Typical logic input connection
Logic output connector
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)