16
Instructions for Use
Use
Switching on and Use of TV 303
To switch on the TV 303 it is necessary to supply the mains. The
integrated controller automatically recognizes start signal presence
and start up the pump.
"Soft Start" mode is used to start the pump after it has stopped for a
long time. To start with the "Soft Start" function on, activate the
mode in the software (see paragraph "RS 232/485 COMMUNICATION
DESCRIPTION" in the "Technical Information" appendix ).
The green LED located on the TV 303 base front panel indi-cates with
its flashing frequency the system operating condi-tions:
TV 303 Switching off
To switch off the TV 303 it is necessary to remove the mains. The
integrated controller immediately stops the pump.
WARNING!
The Turbo-V controller must be powered with 3-wire power cord (see
orderable parts table) and plug (internationally approved) for user's safety.
Use this power cord and plug in conjunction with a properly grounded power
socket to avoid electrical shock and to satisfy CE requirements. High voltage
developed in the controller can cause severe injury or death. Before servicing
the unit, disconnect the input power cable.
Emergency Stop
To immediately stop the TV 303 in an emergency condition it is
necessary to remove the supply cable from the mains plug.
204/268
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.
TV 303 User Manual / 87-901-019-01
(A)