During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
4.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
4.3

Required aids

Reprocessing step Items Needed

Precleaning

Soft cloth
Manual cleaning/
Soft disposable cloth (lint-free)
disinfection
Compressed air gun (2 bar maximum)
4.4
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft cloth. To do so,
place the product in a water bath and/or rinse the product
under running water.