SimplyGo User Manual
Caution: When the SimplyGo device is connected to a CPAP device, you can
only use the Continuous Flow operating mode�
3. Sleep Mode Screen
The Sleep operating mode is similar to the Pulse Mode� The device
detects when the user begins to take a breath using a more sensitive
detection level� The SimplyGo then delivers a pulsed volume of oxygen
determined by the setting number� During inhalation, a low amplitude
and longer pulse duration is used to make sleeping more comfortable� If
no breath is detected for approximately 15 seconds, the system automati-
cally delivers a continuous flow of oxygen at a rate consistent with the
last used Continuous Flow mode setting, or a default of 2�0 LPM�
Selecting the Operating Modes
You can use the Mode button (M) to select the device operating mode�
To select your operating mode:
1�
Press the Mode button�
Note: If the device is operating and the screen is dark, the first press turns
on the screen backlight� In this case, press the button a second time
to select the mode� The intensity of the backlighting can be either
increased or decreased� Refer to the Turning the SimplyGo Device Off
section of this manual for more information�
2�
Continue pressing the Mode button to scroll through the Pulse, Continu-
ous Flow, or Sleep modes�
3�
When your desired mode appears in the display, stop pressing the button
and wait� The display will return to the run time screen of the mode you
selected, and the device will operate in this mode�
18