Philips RESPIRONICS SimplyGo Manual De Usuario página 26

Ocultar thumbs Ver también para RESPIRONICS SimplyGo:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Visual, Audio
Indicators
A sequence of
3 long beeps
+
1 short beep
that repeats every
4 seconds
Description
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering
a lower concentration of
oxygen than specified�
This alarm can be
silenced by pressing the
Alarm symbol�
Note: This alarm occurs
when the internal O
sensor reads < 82% O
This alarm will turn off
if the system is able to
recover from the fault
and achieve >84% O
Note: O
purity is
2
measured at 20 minute
intervals� This alarm
could take a maximum
of 31 minutes to
activate from the time
the alarm condition is
present�
SimplyGo User Manual
What To Do
Change to another
2
source of oxygen
2
and contact your
equipment provider�
2
23

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido