Descargar Imprimir esta página

Safety Standards - Gossen MetraWatt METRACLIP 41 Manual De Instrucciones

Ocultar thumbs Ver también para METRACLIP 41:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

All manuals and user guides at all-guides.com
Unclamp the clamp meter from the conductor, turn it off
using the On - Off switch.
Loosen the captive screw which secures the battery
cover. Lift the cover through 30° and pull it clea r of the
body as shown in Fig 1. Replace the battery and re-fit
the battery cover and fasten the screw .
Fit only Type 9 V PP3 Alkaline (MN 1604).

5. SAFETY STANDARDS

EN 61010-1:2001
EN 61010-2-032:2002
EN 61010-031:2002
300V Cat III Pollution Degree 2
EMC Standards
EN 61326-2-2:2006
ROHS and WEEE compliant
This product is designed to be safe under the following
conditions:
- indoor use
- altitude up to 2000m
- temperature 0° C to +50° C
- maximum relative humidity 80% for
temperatures up to 31° C decreasing linearly
to 40% relative humidity at 50° C.
Use of the current clamp meter on uninsulated
conductors is limited to 300V AC
frequencies below 1kHz.
Safety in its use is the responsibility of the operator
who must be a suitably qualified or authorised person.
Ensure that your fingers are behind the protective
barrier see FIG 1 when using the clamp meter. Always
inspect the clamp meter for damage before use.
To avoid electric shock, keep the clamp meter clean
and free of surface contamination.
Use Isopropyl alcohol to clean the clamp meter.
or d.c. and
RMS
16

Publicidad

loading

Este manual también es adecuado para:

Metraclip 410