4 USING THE LOGGER (continued)
Logging can also be stopped manually by the user at anytime, which
will override any pre-programmed stop triggers.
To manually stop logging:
1
Press the 'Func' key to wake the logger.
2
Press and hold the 'Set' key to stop logging.
When logging has stopped, 'run complete' is displayed and if cure
parameters have been set via ElcoMaster , see Appendix A: Section
A.3 'Creating a New Paint Type' on page en-15, an indication will also
be given whether the latest run has met the cure parameters.
Batch data can be downloaded to ElcoMaster for further reporting
and analysis, see Section 4.4.

4.4 DOWNLOADING DATA

When one or more runs have been recorded, the data can be
downloaded to PC for further analysis and reporting using
®
ElcoMaster
software and the USB cable supplied, see Appendix A
'Using ElcoMaster ' on page en-13 for further information.

4.5 CLEARING THE MEMORY

The Elcometer 215 can store up to 260,000 readings in up to 8 production
runs. When all 8 runs are complete, start logging will automatically
overwrite the oldest. It is therefore not necessary to clear the memory
however, this can be done if required.
To clear the memory:
1
If idle, press the 'Func' key to wake the logger.
2
Repeatedly press the 'Func' key until 'TOOLS' is displayed.
3
Repeatedly press the 'Sel' key until 'clear' is displayed.
4
Press and hold the 'Set' key for approximately 5 seconds whilst
'Hold to continue' is displayed. 'ok' is displayed and all runs are
deleted.
Note: All stored runs will be deleted. It is not possible to select which runs are
deleted.