buttocks. The "pigtails" should hang
down from the pads.
On the front, in a vertical format:
In the vicinity of
your groin, place
the
pads
vertically, about
10cms (4 inches)
apart.
The
"pigtails" should
be uppermost.
On the front, in a diagonal format:
In the vicinity of
your groin, place
the pads in a V-
shape,
narrowing
to about 10cms (4
inches) apart at the
lower end of the
pads. The "pigtails"
should
be
uppermost.
The strength should be comfortable,
but always remain noticeable. You may
need to increase it over the course of
the treatment.
There is no need to have a contraction.
Caution: The self-adhesive
pads supplied are intended for
multi-use by one person only. Do not
lend used pads to others, nor use
someone else's used pads yourself.
16. CLEANING
It is important that the probe is cleaned
after each use. Clean with either an
alcohol-free antibacterial wipe such as
TensCare Wipes (see X-WIPES) or by
wiping with warm soapy water. Rinse
and dry thoroughly and return the unit
to the storage pouch. Do not immerse
the probe in a liquid.
Clean the case of the unit and lead wire
at least once a week using the same
method.
• Do not immerse your Sure+ unit in
water.
• Do not use any other cleaning
solution.
17. EMC
Wireless communications equipment
such
as
wireless
devices,
mobile
telephones and their base stations,
walkie-talkies can affect this equipment
and should be used no closer than 30
cm (12 inches) to any part of the device.
(Note. As indicated in 5.2.1.1(f) of IEC
60601-1-2:2014 for ME EQUIPMENT).
Note: For hospital use, full EMC
advice tables are available on
request.