8. Apply outer layers of material.
9. Perform 1st lamination.
10. Transfer socket alignment:
Optional: If also adding RevoFit® adjustability, install those components now.
11. Perform 2nd lamination.
12. Remove socket from mold.
Remove lamination post from distal housing.
13. Prepare collar:
a. Grind to silicone and remove.
b. Continue grinding carbon until flat with top of the collar.
c. Trim tube.
14. Drill a 3 mm pilot hole into socket frame to access the reel release mechanism.
Note: Drill through lamination into void - Do not damage collar.
15. Lace the device:
a. Use the lace feeder to feed the lace through tubing starting at the collar.
Feeder
Lace
16. Attach lace to the Click® Reel.
8. Aplicar las capas exteriores de material.
9. Realice la primera laminación.
10. Transferir la alineación del zócalo:
Opcional: Si también se añade la posibilidad de ajuste RevoFit®, instale esos componentes ahora.
11. Realice la segunda laminación.
12. Retire el encaje del molde.
Retire el poste de laminación del alojamiento distal.
13. Prepare el cuello:
a. Rectificar hasta la silicona y retirar.
b. Continúe esmerilando el carbono hasta que quede plano con la parte superior del collar.
c. Recorte el tubo.
14. 14. Taladre un agujero piloto de 3 mm en el marco de la toma para acceder al mecanismo de
liberación del carrete.
Nota: Perforar a través de la laminación en el vacío - No dañar el collar.
15. Encaje el dispositivo:
a. Utilice el alimentador de encaje para introducir el encaje a través de la tubería empezando por
el collar.
Alimentador
Encaje
16. Conecte el encaje al carrete Click®.