The instrument can be autoclaved at 121 °C (250 °F), 2 bar with a hold time of at least 15 minutes
according to DIN EN 285.
Preparation for autoclaving
1.
The instrument must be carefully cleaned prior to autoclaving
(see 'Cleaning', page 49-50).
2.
Open the closure cap on the discharge tube, and for instruments
with a recirculation valve, set the valve to 'Dispense'.
3.
Check that the filling valve is securely seated (Fig. 1a). In the
digital model, additionally check that the piston mounting is
securely seated (Fig. 1b).
4.
To ensure unhindered access for the steam and to prevent the
ball valve in the filling valve from possibly becoming stuck, hold
the instrument with the discharge piston pressed vertically down-
ward, and gently tap against the casing with your hand (Fig. 2).
Then lay it horizontally in the autoclave.
Be sure to avoid the instrument coming into contact with metal
surfaces in the autoclave!
Note:
Do not reassemble the instrument until it has cooled down to room
temperature (Cooling time approx. 2 hours).
After every autoclaving, inspect all parts for deformities or damage. If
necessary, replace them.
It is the user's responsibility to ensure effective autoclaving.

Autoclaving

1a
1b
2
53