Descargar Imprimir esta página

Minebea Intec IUXS4 Serie Instrucciones De Instalación Y Operación página 18

Ocultar thumbs Ver también para IUXS4 Serie:

Publicidad

Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
T155°C IP6X
1
For gas
Vi
13 V
Ii
483 mA
Pi
2.0 W
Ci
16nF
Li
6µH
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA 04 ATEX1074 X
2: Standard version (no temperature sensor connected)
6-wire cable installed flexibly and connected to weighing
platform.
3: All metal parts must be grounded (earthed).
Weighing platform
IUXS4-......-.....
or
IFXS4-......-.....
KEMA 04ATEX1074 X
II 2G 1D
EEx ib IIC T4
T155°C IP6X
1
For gas
Vi
13 V
Ii
483 mA
Pi
2.0 W
Ci
16nF
Li
6µH
Remarks
1: Parameters as listed in the EC type examination certificate
KEMA04 ATEX1074 X
2: All metal parts must be grounded (earthed).
18
Operating Instructions IUXS4
Sensor cable connected to the weighing platform
(included in std. equipment supplied): maximum
length 50m. Shields ("screens") installed on both
ends
1
For dust
13 V
425 mA
2.0W
---
---
Hazardous area
Gas:
Zones 1, 2
IIC T4; -20°C ... +40°C
Dust: Zones 20, 21, 22
T155°C; -20°C ... +40°C
Written by
Reviewed by
Released by
Sensor cable connected to the weighing platform
(included in std. equipment supplied): maximum length
50m. Shields ("screens") installed on both ends
1
For dust
13 V
425 mA
2,0W
---
---
Hazardous area
Gas:
Zones 1, 2
IIC T4; -20°C ... +40°C
Written by
Reviewed by
Released by
2
Sensor Output
Vo
12.6 V
Io
350 mA
Po
927 mW
Co
1.15 µF
Lo
220 µH
Lo/Ro
32 µH/
Connection to CIXS3
Date
Name
15.02.06
Klausgrete
Verification of Intrinsic Safety
15.02.06
Klausgrete
15.02.06
Klausgrete
Circuit:
Sensor Output
Vo
12.6 V
Io
304 mA
Po
1.96 W (FCT01-X)
1.34 W (FCT01-XV1)
Co
400 nF
Lo
230 µH
Connection to FCT01-X..
Date
Name
15.02.06
Klausgrete
15.02.06
Klausgrete
15.02.06
Klausgrete
Indicator
CIXS3
KEMA 03ATEX1157 X
II 2 G EEx ib IIC T4
II 1 D IP6X T135°C
-20°C ... + 40°C
For further connections
(power supplies, data
interfaces, etc.), see
65607-099-60-A4
1
Page
5
of
Rev.
36116-741-60-A4
01
Indicator
FCT01-X..
KEMA 00ATEX1012 X
II 2 G
EEx ib IIC T4
-20°C ... + 40°C
For further connections
(power supplies, data
interfaces, etc.), see
65638-099-60-A4
2
Page
5
of
Verification of Intrinsic Safety
Rev.
36116-741-60-A4
01

Publicidad

loading

Este manual también es adecuado para:

Iuxs4-ce serie