13. Compatible accessories
Danger!
The use of accessories and combination of products not authorized for the prod-
uct can lead to serious hazards. This is why only accessories released by wiss-
ner-bosserhoff GmbH and products released for combination may be used.
The compatibility of the accessories listed here may vary depending on the model.
13.1 Mattresses
Only use the bed with a suitable compatible mattress.
Caution!
Do not use water mattresses. If the safe working load is exceeded, the drives and
other components may fail and cause accidents. Never exceed the safe working
load.
Caution!
Only replace wissner-bosserhoff mattresses with equivalent mattresses of the
same dimensions. Make sure that only mattresses with incisions on the underside
that allow for movement are used, as a mattress without these could affect the
adjustment functions or overload the drives. Do not use spring mattresses.
We recommend to use mattresses manufactured with a minimum volume weight of 40 kg/
m³ (density 40 kg/m³). All original wissner-bosserhoff standard mattresses fulfill this require-
ment. Please refer to the following overview for the suitable mattress dimensions for the cor-
responding patient surfaces.
Model
Beds with standard patient
surface:
movita, movita casa, movita 2-d
Beds with narrow patient surface:
movita 2-s
Beds with narrow and short pa-
tient surface: movita 2-xs
100
Mattress dimensions
Aero/Easyclean patient surface 12-18 cm x 87 cm x 200 cm
Comfort patient surface 12-15 cm x 87 cm x 200 cm
Aero/Easyclean patient surface 12-18 cm x 80 cm x 200 cm
Comfort patient surface 12-15 cm x 80 cm x 200 cm
Aero/Easyclean patient surface 12-18 cm x 80 cm x 190 cm
Comfort patient surface 12-15 cm x 80 cm x 190 cm