Table 2 Guidance and MANUFACTURER's declaration – electromagnetic
IMMUNITY – for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE
U
is the a.c. mains voltage prior to application of the test le vel.
T
IEC 60601-1-2
Test level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.0 % UT;
1 cycle and 70 % UT; 25/30 cycles;
Single phase: at 0°.0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MH z
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MH z
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance level
±8 kV contact
±2 kV, ±4kV, ±8 kV,
±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MH z
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MH z
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
28