Hach NA5600 sc Na+ Mantenimiento Y Solución De Problemas página 12

Ocultar thumbs Ver también para NA5600 sc Na+:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 119
Prepare the analyzer for a long-term shutdown (more than 1 week) as follows:
Items supplied by user:
• Small container, 250 mL or larger
• Deionized water, 1.5 L
• Clean, no-lint cloth
Items to collect:
• Storage bottles for electrodes (2x)
• Protective caps for electrodes (2x)
• Black cover for glass junction of reference electrode
3.9.1 Drain and rinse the KCl electrolyte reservoir
Refer to the illustrated steps that follow.
Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
12 English

Publicidad

Tabla de contenido
loading

Tabla de contenido