Factory Calibration Reset - Aqualytic MD110 Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
All manuals and user guides at all-guides.com
Calibration Mode
GB

Factory calibration reset

Resetting the user calibration to the original factory calibration will
reset all methods and ranges.
Store
Date
A user calibrated method is indicated by an arrow while the test result
Cal
RESULT
is displayed.
Time
Cal
To reset the calibration press both the [MODE] and [ZERO/TEST] key
Zero
Mode
and hold.
Test
Switch the unit on using the [ON/OFF] key.
On
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
Off
The following messages will appear in turn on the display:
SEL
The factory setting is active.
(SEL stands for Select)
CAL
or:
Calibration has been set by the user.
SEL
(If the user calibration is to be retained, switch the unit off using the
cAL
[ON/OFF] key).
Mode
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
SEL
CAL
On
Switch the unit off using the [ON/OFF] key.
Off
AL110_2 03/2018
25

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Al 110

Tabla de contenido