3.2 Laser Specification
• Laser Safety Classification: Class II.
• Wavelength: Red (630 ~ 670nm).
• Output Power: < 1mW (Class II).
3.3 Electrical Specification
• Temperature Range: -30°C ~ 550°C / -22°F ~ 1022°F.
• Display Resolution: 0.5 / 1°C automatic, 1°F.
• Accuracy:
± (2°C / 4°F) to -30°C ~ 100°C (-22°F ~ 212°F).
± (2% reading) to 101°C ~ 550°C (213°F ~ 1022°F).
• Temperature Coefficient:
± 0.2% of reading or ± 0.2°C / 0.36°F, whichever
is greater, change in accuracy per °C or °F change
in environment operating temperature above 28°C
/ 82.4°F or below 18°C / 64.4°F.
• Response Time: 250ms.
• Spectral Response: 6 to 14µm.
• Adjustable Emissivity: 0.1 ~ 1.0.
• Detection Element: Thermopile.
• Optical Lens: Fresnel Lens.
• Sighting: 1 Beam Laser Marker < 1mW (Class II).
• Field of View: 100mmØ at 1000mm of distance
(10:1).
4