package, labworldsoft, provides a convenient tool for control-
ling stirring equipment and collecting data under MS Windows,
and includes graphical entry features, for motor speed ramps
for example.
The following table summarises the (NAMUR) commands
understood by the IKA control equipment. Abbreviations used:
X=4:speed X=5:torque
Ramp commands
The IKA stirrer appliances with integrated serial interface offer
the great advantage in ramp mode that the speed ramps which
the stirrer machine is supposed to work off are stored in the
stirrer machine and not in the PC.
Through the IKA specific RMP_LOOP_SET command there is
even the possibility of transmitting an endless ramp to the stir-
rer appliance.
NAMUR
Commands
IN_PV_X
X = 4; 5
OUT_SP_X n
X = 4
IN_SP_X
X = 4
START_X
X = 4
STOP_X
X = 4
RESET
STATUS_X
EUSTAR0707
Function
Read actual value
Set actual value to n
(Maximum value: 1999)
Read actual value input
Switch on appliance (remote) function;
(indication: remote)
Switch off appliance function. The variable
set with OUT_SP_X remains. Includes the
command RMP_STOP. (Indication:remote)
Switch to normal operation. Only possible
if the speed control is set to n < 60.
Status output
0:manuel operation without interruption
1:automatic operation Start (without inter
ruption)
2::automatic operation Start (without inter
ruption)
<0: fault code :
-83: parity fault
-84: unknown command
-85: incorrect command sequence
-86: invalid nominal value
-87: insufficient memory
RMP_IN_X
Read current segment number of ramp.
X = 4
If ramp no started: 0
RMP_IN_X_y
Read end value and ramp segment dura-
tion for ramp segment y.
RMP_OUT_X_y n Set end value (n) and ramp segment dura-
hh:mm:ss X = 4
tion hh:mm:ss for ramp segment y.
RMP_START_X
Switch on ramp function starting with ramp
X = 4
segment no.1. Only possible after previous
START_X. After RMP_STOP_X, however,
START_X is not necessary.
RMP_STOP_X
Switch off ramp function. Nominal value
X = 4
= 0. (Ramp remains, ramp can be started
again with RMP_START_X).
RMP_PAUSE_X
Halt ramp function. Freeze current nominal
X = 4
value and current ramp segment time.
RMP_CONT_X
Continue ramp function. (After previous
X = 4
RMP_PAUSE_X)
RMP_RESET_X
Switch off ramp function and cancel all
previously input ramp segments.
Further commands
IKA specific
RMP_LOOP_SET_X Work off ramps in a loop
X = 4
RMP_LOOP_RESET_X End ramp loop
X = 4
IN_TYPE
Request laboratory appliance identification
IN_NAME
Request designation
OUT_NAME name
Output designation name (max. 6 charac-
ters, default: IKA_RW)