5.4
Recommended methods
Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
5.5
Dismantling
1. Remove the HF electrode from the handle.
2. For conization electrode Art. No. 20191-075 only: Remove
the loop out of the conization electrode.
5.6
Required Aids
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
5.7
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
19