Usb Mode; Mass Storage; Com Interface - Lovibond SD 335 Multi Manual Del Usuario

Ocultar thumbs Ver también para SD 335 Multi:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 95

4.1.2 USB mode

4.1.2.1 Mass storage

When the USB Mass Storage mode is selected, the machine cannot access the internal memory.
The logger can then no longer be started. In this mode, you can access the memory directly with-
out a driver and copy the stored measurement data to the computer or delete them from the
memory. The measurement data are available as a csv file. The storage location is linked to the
start of the logger, e.g. 31st December 2020 19:11 of the associated files can then be found in the
folder \DATA\20201231\1911\. Single value recordings are always stored in the \HISTORY folder.
Calibration data are stored like logger data but in the folder \CAL_DATA

4.1.2.2 COM interface

In this mode the data logger can be operated. Furthermore, after driver installation (only
Windows drivers available), communication with the device is possible. (115200 8N1 \r\n as
end recognition).
The following commands are supported:
GetChannelMenu: #
GetLastValue: #
GetCalibrationReport: #
GetDeviceInformation: 0
AddLocationDescription: ## Text Changes the description from location ## to text
# corresponds to the channel number starting at 0
No.
Parameter
0
Oxygen
1
pH
2
Conductivity
3
Temperature (of pH input)
4
Temperature (from O
5
Air pressure
6
Device menu (measured value = battery level in %
44
Output of all parameter settings
Output of the last measured value
Output of the last calibration data
Devices and License Information
/Con input)
2
Filename (Prefix)
O
2
PH
COND
T_PH
T_COND
PRES
DEV
SD_335_1 10/2020

Publicidad

Tabla de contenido
loading

Tabla de contenido