Proservo NMS81
Endress+Hauser
11.4
Visión general de los mensajes de diagnóstico
Número de
Texto corto
diagnóstico
Diagnóstico del sensor
102
Sensor incompatible error
150
Detector error
151
Sensor electronic failure
Diagnóstico de la electrónica
242
Software incompatible
252
Modules incompatible
261
Electronic modules
262
Module connection
270
Main electronic failure
271
Main electronic failure
272
Main electronic failure
272
Main electronic failure
273
Main electronic failure
275
I/O module failure
276
I/O module faulty
282
Data storage
283
Memory content
284
Detector SW update in
progress
311
Electronic failure
Diagnóstico y localización y resolución de fallos
Remedio
1. Restart device
2. Contact service
1. Restart device
2. Check electrical
connections of detector
3. Replace detector unit
Replace sensor electronic
module
1. Check software
2. Flash or change main
electronics module
1. Check if correct electronic
modul is plugged
2. Replace electronic module
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
1. Check module connections
2. Change electronic modules
Replace main electronics
1. Restart device
2. Change main electronic
module
Restart device
1. Restart device
2. Contact service
1. Emergency operation via
display
2. Change main electronics
1. Restart device
2. Change I/O module
1. Restart device
2. Change I/O module
1. Restart device
2. Contact service
1. Transfer data or reset
device
2. Contact service
Firmware update active,
please wait!
Maintenance required!
1. Do not perform reset
2. Contact service
Señal de
Comportamiento
estado
de diagnóstico
[Ex-
[Ex-fábrica]
fábrica]
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
M
Warning
145