Manual Cleaning/Disinfecting; Manual Cleaning And Wipe Disinfection - B.Braun Acculan 3Ti Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 78

9.5 Manual cleaning/disinfecting

Manual cleaning and wipe disinfection

Stage
Step
I
Cleaning
II
Drying
III
Wipe disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized)
RT:
Room Temperature
Stage I
Clean the product under running tap water, using
a suitable cleaning brush if necessary, until all
visible residues have been removed from the
surfaces.
Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
Make certain the product is positioned in such a
way that water will not enter the product e.g.
through coupling interfaces. (Immediately remove
any fluid that entered inadvertently.)
Use the special cleaning brush (TA011944) for
cleaning the cannulation of drill and reamer
attachments/Kirschner wire attachment GB643R.
To avoid the risk of corrosion, do not use a metal
brush or other abrasives that would damage the
surfaces for cleaning the product.
Clean difficult-to-access areas with soft round
plastic brushes of fitting diameter.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
-
-
(cold)
RT
-
-
≥1
-
-
RT
0.5
-
(cold)
RT
-
-
Stage II
Stage III
Stage IV
Water
quality
D–W
-
-
-
-
Meliseptol HBV wipes
50 % Propan-1-ol
FD–W
-
-
-
Dry the product with lint-free tissue or medical-
quality filtered compressed air.
Wipe all surfaces of the product with a single-use
disinfecting wipe.
After the specified exposure time (of at least
1 minute), rinse the disinfected surfaces under
running FD water.
Hold the product in such a way that water cannot
enter it e.g. through couplings or adapters.
Allow water to drip off for a sufficient length of
time.
Chemical
17

Publicidad

Tabla de contenido
loading

Tabla de contenido