10.2 Malfunctions of the Kirschner wire attachment
Malfunction
Kirschner wire cannot
be inserted into
Kirschner wire
attachment
Kirschner wire not
turning
10.3 Malfunctions of the sagittal saw attachment
Malfunction
Saw blade cannot be
uncoupled
Cause
Incorrect adjustment of
Kirschner wire
attachment
Incorrect adjustment of
Kirschner wire
attachment
Clamping lever open
Cause
Release button not
pressed through
properly
Finding
Kirschner wire diameter
not matching the
setting of the adjustment
sleeve
Kirschner wire diameter
not matching the
setting on the
adjustment scale
Clamping lever open
Finding
Locking pin not fully
releasing the saw blade
Remedy
Set the adjustment sleeve to
the correct Kirschner wire
diameter
Set the adjustment sleeve to
the correct Kirschner wire
diameter
Close the clamping lever
Remedy
Fully press down the release
button, see Coupling the saw
blade
23