Installation
1
Installation
1.1
Range of application
1.2
Installation, Setup
Rack mounting
2
The CenterOne, CenterTwo or CenterThree devices are to be installed and operated in
the following ambient conditions:
Installation location
Degree of protection
Installation altitude
Ambient temperature
Relative humidity
Air pressure
The unit can be operated in three different ways:
– mounted in a 19" rack cabinet
– mounted in an operator panel
– as a desktop unit
Commissioning of a defective product
Putting a product which presents a visible damage into operation can be extremely haz-
ardous.
If the product presents a visible damage do not put it into operation and make sure it
is not inadvertently put into operation.
The unit is designed for installation into a 19" rack chassis adapter according to DIN 41
494. For this purpose, four collar screws and plastic sleeves are supplied with it.
Loss of the protection class of the rack
If the product is installed in a rack, it is likely to lower the protection class of the rack
(protection against foreign bodies and water) e.g. according to the EN 60204-1 regula-
tions for switching cabinets.
Take appropriate measures for the rack to meet the specifications of the protection
class.
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!
sheltered (interior)
IP 30
max. 2000 m
+5°C to +50°C
≤ 80% up to +31°C, decreasing to 50% at +40°C
860 hPa - 1060 hPa
DANGER
NOTICE
NOTICE