Installation
Pin assignment
CH 1, CH 2 and CH 3
8
Pay attention to the list of usable measuring tubes!
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Multiple assignment
Possibility of damage if transmitters are connected simultaneously.
Only connect one transmitter to each measuring channel (terminal CH A or CH B).
i
Fig. 9:
Only one transmitter per measuring channel
RJ45 socket
The pins on the 8-pin RJ45 socket (FCC68) are assigned as follows:
Pin
Signal
1
Infeed (+24 V DC)
2
Infeed ground (GND)
3
Signal input (measuring signal 0 ... +10 V DC)
4
Identification
5
Signal ground
6
Status
7
HV_L
8
HV_H / HV_EMI
Fig. 10: RJ45 socket
Sub-D socket
The pins on the 15-pin sub-D socket are assigned as follows:
Pin
Signal
1
EMI status
2
Signal input (measuring signal 0 ... +10 V DC)
DANGER
NOTICE