Error Messages/Trouble-Shooting - Beurer BM 58 Manual De Instrucciones

Ocultar thumbs Ver también para BM 58:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 53
PC is shown on the display. Begin the data transfer
in the "beurer HealthManager" PC software.
During the data transfer, an animation is shown
on the display. A successful data transfer is
displayed as in figure 1. If the data transfer is
unsuccessful, an error message appears as in
figure 2. In this case, interrupt the PC connection
and start the data transfer again.
After 30 seconds of not being in use or if com-
munication with the PC is interrupted, the blood
pressure monitor switches itself off automati-
cally.

9. Error messages/trouble-shooting

In case of faults, the
_ message appears in the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured
(
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely
(
or
appears on the display)
• the pump pressure is higher than 300 mmHg
(
appears on the display)
All manuals and user guides at all-guides.com
• pumping up takes longer than 160 seconds
• there is a system or device error
• the batteries are almost empty
In the above cases, you must repeat the measurement.
Abb. 1
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace them.
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical speci-
Abb. 2
fications", the technical alarm will appear on the display
indicating either " " or "Lo". In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.
32
(
appears on the display)
(
,
,
or
appears on the display)
Technical alarm – description
.

Publicidad

Tabla de contenido
loading

Tabla de contenido