Step 2.
Press and hold the
displayed for approximately two seconds. Then
release the key.
Step 3.
The balance displays
If you want to change the sensitivity adjustment
weight, press the
4. Refer to "8. CALIBRATION" in the instruction
manual for sensitivity adjustment weights that can
be selected.
If you use the sensitivity adjustment weight value
stored in the balance, proceed to step 5.
Step 4.
Specify the weight value as follows :
key ...........
SAMPLE
key (+) .
RE-ZERO
key (-)
MODE
key ................
PRINT
key .....................
CAL
Step 5.
Confirm that there is nothing on the pan and press the
key. The balance measures the zero point. Do not allow vibration
or drafts to affect the balance. The balance displays the sensitivity
adjustment weight value.
Place the displayed sensitivity adjustment weight on the pan
Step 6.
and press the
adjustment weight. Do not allow vibration or drafts to affect the
balance.
Step 7.
Remove the weight from the pan after displaying
Step 8.
If the GLP output is set, "Calibration Report" is outputted.
Step 9.
The balance will automatically return to the weighing mode.
Step 10. Place the sensitivity adjustment weight on the pan and confirm that the
measurement value is within ±2 digits. If it is out of that range, check the ambient
conditions and repeat from steps 2.
key until
CAL
key and proceed to step
SAMPLE
All segments blink : Weight selection mode.
The last three digits blink : Weight value adjustment.
All segments blink : Select weight.
The last three digits blink : Weight value adjustment.
Weight value is stored in memory. Data is maintained
even if removing power supply.
Operation is canceled and display returns to
key. The balance measures the sensitivity
PRINT
is
.
6
Press and hold for
approx. 2 sec.
New value example
Zero point
Sensitivity
.
adjustment weight
PRINT.
GLP output
.