5.
Maintenance
5.1
Maintenance intervals
i
Routine maintenance ensures constant readiness for
operation.
The following maintenance intervals are to be observed
unless otherwise specified by the approval authorities.
5.1.1
Half-yearly maintenance
¶
Replacement of coarse filter GF1 (Fig. 4, Pos. 8) in
sampling hose.
¶
Replacement of coarse filter GF2 (Fig. 4, Pos. 1),
refer to Section 5.2.4.
¶
Check whether both PVC hoses are connected to the
gas outlets (Fig. 4, Pos. 6 and 7).
¶
Visual inspection of sampling probe (Fig. 4, Pos. 10).
¶
Sampling probe leak test (refer to Section 5.2.1).
5.1.2
Annual maintenance
This work is to be performed by a qualified maintenan-
ce service. It includes the half-yearly maintenance and
the following additional items:
¶
Checking of BEA 060 measurement accuracy with
a test gas.
¶
Replacement of activated charcoal filter
(Fig. 4, Pos. 4) in zero gas path.
¶
Replacement of coarse filter GF3
(refer to Section 5.2.4)
Compliance with the specifications of the relevant
authorities must be ensured.
Robert Bosch GmbH
All manuals and user guides at all-guides.com
Fig. 3:
1 Coarse filter GF2
2 Coarse filter GF3
3 8 m hose
4 Activated charcoal filter AF1
5 O
6 Measurement gas outlet (transparent PVC hose)
7 Gas outlet and condensate outlet (transparent PVC hose)
8 Coarse filter GF1
9 30 cm Viton hose (black)
10 Exhaust gas sampling probe
11 Plastic hose for leak test
5.2
5.2.1
To ensure accurate exhaust gas measurements it is es-
sential that there are no leaks in the sampling system
(exhaust gas sampling probe, hoses, coarse filters). It is
therefore advisable to conduct the leak test every day.
5.2.2
Keep the opening at the tip of the exhaust gas sampling
probe clean. In the event of HC residue and conden-
sate, detach the exhaust gas sampling probe from the
hose and blow compressed air through it in the direc-
tion opposite to intake direction.
5.2.�
Check for damage. In the event of HC residue or con-
densate, detach the hose from the analyzer and blow
compressed air through it in the direction opposite to
intake direction.
Maintenance | BEA ��� | 25
N0
BDM 300
11-16V
20 W
459907_4Ko.eps
7
11
BEA 060
sensor
2
Maintenance
Freedom from leaks of sampling system
Exhaust gas sampling probe
Sampling hose
en
BEA 060
GF 3
GF 2
O2
AF I
1
2
6
5
4
3
8
9
10
|
1 689 989 112
2012-11-26