Functional description

GB

Operation

On
Switch the unit on using the [ON/OFF] key.
Off
METHOD
The display shows the following:
Select the required test using the [MODE] key.
Mode
The display shows the following:
METHOD
Fill a clean vial with the water sample up to the 10 ml mark, screw
the cap on and place the vial in the sample chamber making sure that
the
Zero
Press the [ZERO/TEST] key.
Test
The "Method" symbol flashes for approx. 8 seconds.
METHOD
The display shows the following:
0.0.0
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
Zero
Press the [ZERO/TEST] key.
Test
The "Method" symbol flashes for approx. 3 seconds.
METHOD
The result appears in the display.
RESULT
The result is saved automatically.
Repeating the test:
Zero
Press the [ZERO/TEST] key again.
Test
Repeating the zero:
Zero
Press the
Test
Caution:
A calibration of the method has to be performed before the first use
(see
of SPADNS reagent solution must be used for adjustment and test.
The adjustment process needs to be performed for each new batch of
SPADNS reagent solution.
20
All manuals and user guides at all-guides.com
marks are aligned.
marks are aligned.
[ZERO/TEST]
key for 2 seconds.
4
"menu-options"
calibration mode Fluoride). The same batch
MD100_5 10/2014