OPERATION GUIDELINES
1
UNIT SET UP
1. Connect the power cord to the back of the unit
and to an outlet.
2. Turn the power switch found on the back of the
unit to the On position.
3. LED display shows 10:00. This is the most
common timer setting.
If the timer needs to be adjusted,touch TIME+ key for
quick timer settings and Time– key for fine adjustment
between 1 to 30 minutes.
Please consult your instrument cleaner instructions as
well as your instrument directions for use for proper
automated cleaning.
Prolonged cleaning time may result in:
1. Loosening of the screws if used.
2. If items are plated and have existing cracks, the
cracks may become more pronounced.
3. Coatings with preexisting damages may
experience more peeling.
2
DEGAS PROCEDURE
1. Remove the cover and put it on the cover holder
vertically. Add room temperature water to below
the MAX but higher than the MIN fill lines.
2. Add OPTIM 28 Instrument Cleaning Solution per
its Instructions for Use.
3. Touch the Degas key and allow it to operate for
90 seconds.
Please consult your Instrument cleaning Solution
Instructions for Use to determine if degassing is
required. OPTIM 28 does not require degassing.
LOADING INSTRUMENTS
3
1. Put items in cassettes or basket and lower into
the cleaning tank.
2. Confirm that the water level is between MIN and
MAX and above the area to be cleaned.
3. If not, add room temperature water to the tank.
If the unit is turned on without water, ultrasonic
energy will not be absorbed. Due to this
unit's high power, if on without water for over
15 seconds damage to the unit may occur,
severely reducing the life of the unit.
5