ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used in
such environments:
Immunity test
IEC 60601 test level
Electrostatic
± 8 kV contact
discharge (ESD)
± 15 kV air
IEC 61000-4-2
Electrical fast
power supply lines: ± 2 kV
transient/burst
input/output lines: ± 1 kV
IEC 61000-4-4
Surge
line(s) to line(s): ± 1 kV
IEC 61000-4-5
line(s) to earth: ± 2 kV
100 kHz repetition
frequency
Voltage
0% UT; 0.5 cycle
dips, short
At 0°, 45° , 90° , 135° , 180° ,
interruptions and
225° , 270° and 315° .
voltage variations
0% UT; 1 cycle
on power supply
70% UT; 25/30 cycles
input lines
0% UT; 250/300 cycles
IEC 61000-4-11
Power frequency
30 A/m
(50 or 60 Hz)
50 or 60 Hz
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment – guidance
Group 1
RF energy is used only to maintain device's operation. Therefore, its
RF emissions are so low that it's not likely to cause any interference
Class B
in nearby electronic equipment.
Class A
The device is suitable for use in all establishments, including
domestic establishments, and those directly connected to the public
low-voltage power supply network that supplies buildings used for
Complies
domestic purposes.
Compliance level
± 8
± 15 kV air
power supply
lines: ± 2 kV
line(s) to line(s):
± 1 kV
100 kHz repetition
frequency
0% UT; 0.5 cycle
At 0°, 45° , 90°, 135° ,
180°, 225° , 270°
and 315° .
0% UT; 1 cycle
70% UT; 25 cycles
0% UT; 250 cycles
30 A/m
50 or 60 Hz
Electromagnetic environment –
guidance
kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.