Accessories
Oxygen Outlet Connector - Plastic - 1/pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CN100
Bubble Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Salter Labs 7600 or equivalent
There are many types of humidifiers, oxygen tubing and cannulas/masks that can be used with this device. Certain humidifiers and accessories may impair the device's
performance. A mask or any nasal cannula can be used with continuous flow delivery and may be sized according to your prescription as recommended by your homecare
provider who should also give you advice on the proper usage, maintenance, and cleaning .
NOTE– A maximum of 50 feet (15 meters) of crush-proof oxygen tubing plus 7 feet (2.1 meters) of cannula plus a bubble humidifier is allowed between the concentrator and
the patient.
NOTE– The oxygen supply accessory (patient tubing) shall be equipped with a means that in case of fire stops the delivery of
oxygen to the patient. This means of protection should be located as close to the patient as practicable.
SETTING UP YOUR OXYGEN CONCENTRATOR
1 .
Position your unit near an electrical outlet in the room where you spend most of your time .
WARNING
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing
oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of
flame.
NOTE– Do not connect to an electrical outlet controlled by a wall switch. No other appliances should be plugged into the wall outlet.
2.
Position your unit at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper flow of air in and
out of your oxygen concentrator . The oxygen concentrator should be located so as to avoid pollutants or fumes .
BEFORE OPERATING YOUR OXYGEN CONCENTRATOR
1.
Before operating your unit, always check to be sure the air filter (located on the back of your unit) is clean. Proper cleaning of this filter
is discussed in the Caring For Your Concentrator section on page 6 .
2 .
Attach the appropriate oxygen accessories to the oxygen outlet .
Oxygen Tubing Connection:
a .
Thread the oxygen outlet connector onto the oxygen outlet .
b .
Attach the oxygen tubing directly to the connector (Figure 1) .
Oxygen Tubing Connection With Humidification:
If your physician has prescribed an oxygen humidifier as part of your therapy, follow these steps (If using a prefill, go to step b.):
a.
Fill the humidifier bottle with distilled water. Do not overfill.
b.
Thread the wing nut located on the top of the humidifier bottle to the oxygen outlet so that it is suspended (Figure 2). Make sure it
is securely tightened .
c.
Attach the oxygen tubing directly to the humidifier bottle outlet fitting (Figure 3).
3 .
Your physician has prescribed either a nasal cannula or face mask . In most cases, they are already attached to the oxygen tubing . If
not, follow the manufacturer's instructions for attachment .
4 .
Remove the power cord completely from the line cord strap . Make sure the power switch is in the "Off" position, and insert the plug into
the wall outlet . The unit is double insulated to guard against electric shock .
NOTE– (only 115 volt units) The plug on the DeVilbiss oxygen concentrator has one blade wider than the other. To reduce the risk of electric
shock, this plug is intended to fit in a wall outlet only one way. Do not attempt to defeat this safety feature.
WARNING
Improper use of the power cord and plugs can cause a burn, fire, or other electric shock hazards. Do not use the unit if the power
cord is damaged.
OPERATING YOUR DEVILBISS OXYGEN CONCENTRATOR
WARNING
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person
utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or
naked sources of flame.
The DeVilbiss oxygen concentrators manufactured since July 1, 2012 are equipped with a fire mitigating outlet fitting that
prevents propagation of fire into the unit. DeVilbiss oxygen concentrators manufactured prior to July 1, 2012 must be
retrofitted by your provider with the fire mitigating outlet fitting.
In order to prevent a fire propagating from the patient through the cannula towards the unit, a means of protection should be located as close to the patient as
practicable. Please contact your provider for this means of protection.
1 .
Press the power switch to the "On" position . When the unit is turned "On," all four lights (Power, Service Required, Low Oxygen, and Normal Oxygen) on the front panel
will illuminate briefly and audible signal will briefly alarm. After a few seconds, only the Power and Normal Oxygen lights will remain lit.
NOTE– DeVilbiss recommends for optimal service life that the DeVilbiss Oxygen Concentrator to be operated for at least 30 minutes after it is powered on. Shorter periods of
operation may affect the long term reliable operation of the product.
WARNING
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned on but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
NOTE– If the Service Required light illuminates and the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Minor Troubleshooting
chart on page 7 and contact your DeVilbiss provider if necessary.
A-525D3
1
2
3
NO SMOKING
EN - 5