NOTE– If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Minor Troubleshooting chart on page 7, and contact your
DeVilbiss provider if necessary.
2.
Check the flow meter to make sure that the flow meter ball is centered on the line next to the prescribed number of your flow rate.
CAUTION– It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
NOTE– Your DeVilbiss provider may have preset the flow meter so that it can not be adjusted.
NOTE– If the flow meter knob is turned clockwise, the flow decreases (and eventually will shut off the oxygen flow). If the knob is turned
counterclockwise, the flow increases.
NOTE– For prescriptions of 5 LPM, be sure the ball is centered on the 5 liter line; the ball should not touch the red line. Setting the flow
higher than 5 may cause the oxygen purity level to drop.
NOTE– The low-flow alarm will activate if the flow meter ball is set below .3 lpm. The unit will continue to run; however, the Service
Required light will come on accompanied by an audible alarm. Adjust the flow meter to your prescribed flow.
NOTE– Installation of 515LF-607 low output flow meter package will cause the low flow alarm to not work.
3 .
Your DeVilbiss concentrator is now ready for use, properly position the cannula or mask (Figure 4) . Allow 20 minutes for oxygen
concentrator to reach stated performance .
DeVilbiss OSD Operation
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit . When the unit is turned "On," all four lights (Power,
Service Required, Low Oxygen, and Normal Oxygen) on the front panel will illuminate briefly. After a few seconds, only the Power and Normal Oxygen lights will remain lit.
The OSD lights on the front panel are defined as follows:
Green Normal Oxygen light–acceptable oxygen level.
Yellow Low Oxygen light– below an acceptable oxygen level.
If the oxygen purity falls below the acceptable level, the green Normal Oxygen light will shut off and the yellow Low Oxygen light will illuminate . Switch to your reserve oxygen
system . Refer to the Minor Troubleshooting section in this guide on page 7 and contact your DeVilbiss provider .
As an added safety feature should the oxygen purity continue to drop, an intermittent audible signal will sound . Contact your DeVilbiss provider immediately . Do not attempt
any other maintenance .

RESERVE OXYGEN SYSTEM

As a precaution, your DeVilbiss provider may supply you with a reserve oxygen system . If your unit loses electrical power or fails to operate correctly, the Patient Alert System
will sound to signal you to switch to your reserve oxygen system (if provided) and contact your DeVilbiss provider .
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR
DeVilbiss recommends using only original DeVilbiss parts and filters in order to guarantee a reliable operation of the product.
NOTE– Use no lubricants, oils or grease.
WARNING
Before attempting any cleaning procedures, turn the unit "Off."
Cannula/Mask, Tubing, and Humidifier Bottle
Clean and replace the cannula/mask, tubing, and humidifier bottle according to the manufacturer's instructions.
Air Filter and Oxygen Outlet Connector
The air filter and connector should be cleaned at least once a week. To clean, follow these steps:
1.
Remove the air filter, located on the back of the unit. Remove the oxygen outlet connector (if used).
2 .
Wash in a solution of warm water and dishwashing detergent (Figure 5) .
3.
Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
CAUTION– To prevent product damage, do not attempt to operate the unit without the air filter or while the filter is still damp.
Exterior Cabinet
As needed, clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild household cleaner and wipe it dry .
WARNING
To avoid electric shock, do not remove the concentrator cabinet. The cabinet should only be removed by a qualified DeVilbiss technician. Do not apply liquid
directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage.

WARRANTY

DeVilbiss provides warranty for this device . Use only DeVilbiss approved original spare parts for maintenance and repair to retain warranty . Using unapproved or non-original
spare parts would void warranty .
PROVIDER'S NOTES - Cleaning and Disinfection When There is a Patient Change
When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and
the next patient, user or third party should be protected by appropriate handling and preparation .
Therefore, when there is a patient change, people must be protected during the transport and handling of the device, and the device must be fully processed, i .e ., cleaned and
disinfected, by suitably trained personnel before reuse to protect the next patient. The complete processing may only be done by the manufacturer or by a qualified DeVilbiss
provider/service technician .
NOTE–If the following described complete processing of the concentrator by a qualified DeVilbiss provider/technician is not possible, the device must not be used by another