Error Codes List

When internal and some external problems occur with the balance, the internal software
will display messages as error codes. Explanations are given for possible problems.
Try remedies in order indicated.
Error Codes
Err 1 chEct
Err 2 LoAd
Err 4 WEIght Incorrect calibration weight
Err 5 999999 Displayed value >99999
Err 6 count
Err 8 FULL
Err 9 dAtA
Invalid checksum data
a) Cycle balance ON/OFF,
b) Return balance for servicing.
Overload (>cap+90d) or Underload
a) Remove load,
b) Check shipping lock,
c) Recalibrate.
a) Use correct calibration weights,
b) Check environmental conditions.
a) Exit application,
b) Rezero balance.
Parts counting error - balance exits parts counting
a) Exit application,
b) Rezero balance.
RS232 buffer is full
a) Cycle ON/OFF,
b) Check RS-232 settings in balance and computer.
Internal data error. The next button press or a 5 second time-out
causes the balance to turn off.
a) Cycle ON/OFF,
b) Return balance for servicing.