5. INSTALLATION

5.1 SETTING UP THE LIGHT SOURCE
Place the light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with
liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the light source.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
CAUTION Use only cords provided with the light source.
Plug the AC power cord into a wall outlet using the three (3) prong plug supplied with the
unit.
CAUTION
NOTE
5.2 CONNECTING THE LIGHT CABLE
Connect the light cable to endoscope and then plug the light cable endfitting into the
appropriate port of the turret on the front panel.
LIT093CUDA
(English)
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power
with three prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
1 2
3 4
REV. C