Enter the
selection
CURRENT pushbutton, and the display shows:
F R O N T
/
R E M O T E
S E L E C T
I O N : X X X X X X
where: XXXXXX = means the word FRONT or
REMOTE, or RS 232 depending on the last
selection.
The controller is factory-set for FRONT panel
operation.
•
Choose the desired selection by pressing the
CYCLE pushbutton:
F R O N T
/
R E M O T E
S E L E C T
I O N :
by pressing
the
PUMP
/
2 3 2
/
2 3 2
F R O N T
•
Press the PUMP CURRENT pushbutton to enter the
value and the display shows:
R S 2 3 2
B A U D
R A T E
where: XXXX = means 600, 1200, 2400, 4800,
9600 baud rate for the host computer or printer
communication. The controller is factory-set for
9600 baud rate operation.
•
Select the desired value by pressing the CYCLE
NUMBER:
R S 2 3 2
B A U D
R A T E
61
TECHNICAL INFORMATION
X X X X
9 6 0 0