automatically turn off in about 20 seconds. Pressing any button will remove the message from the display.
NOTE–If the DeVilbiss SleepCube does not turn off automatically, the Auto-OFF feature may have been disabled. In this
case you must manually turn the device off.

Enable Menu List

The Enable menu is available in all modes and is a sub-menu that allows you to control various functions of your
SleepCube. It is accessed by pressing the down button while "Enable Menu" is shown on the Display.
Display Example
Auto-OFF
Enabled
Auto-ON
Enabled
Mask Fit Check
Enabled
Low Backlight
Enabled
Exit Menu
Press 

PATIENT MESSAGES

Your SleepCube will alert you of issues that may require some action on your part in order to make your therapy more
effective. You can press any key or button to clear the message from the display, but the SleepCube will operate normally
even if the message is not cleared. Two kinds of messages are displayed: Reminders and Notifi cations.

Reminders

Reminders alert you when parts of your system need to be replaced. Some components of your system wear over time
and, if not replaced, may compromise your therapy. If you see a message on the display that is not listed below, you may
have the optional DeVilbiss SmartLink Module attached to your device (Fig. F), which provides additional messages. Refer
to the documentation that was provided with the SmartLink Module.
Clean Filter – The fi lter should be checked every 10 days for signs of dirt or wear so that it can be cleaned as needed.
This message helps to remind you to check it regularly.

Notifications

Notifi cations identify conditions in your SleepCube that may require action by you or your equipment provider in order to
maintain a high level of therapy. If you see a message on the display that is not listed below, you may have the optional
DeVilbiss SmartLink Module attached to your device (Fig. F), which provides additional messages. Refer to the
documentation that was provided with the SmartLink Module.
Delay Running xx Minutes Left – While a comfort delay session is active, this message is fl ashed on the display every 5
seconds to let you know how much time is left in the delay.
Mask Leak – This message means the SleepCube has detected a large amount of airfl ow during therapy for at least 10%
of the time during your previous use session. This notifi cation is displayed when the device is turned on. If this message is
displayed, put on the mask and adjust the headgear to ensure the mask is properly fi tted to your face. Follow the mask
manufacturer's directions for adjusting the mask and headgear for proper fi tting.
Mask Off – This message appears due to a poor mask fi t or a removed mask. Check for leak around the mask seal and
make adjustments as necessary according to the mask manufacturer's instructions. This message will be displayed until
the high airfl ow problem is corrected. If the high airfl ow condition persists for about 20 seconds, the device will
automatically turn off if Auto-OFF is enabled.
Device Fault – If a device fault message is displayed, refer to Troubleshooting for instructions.
SE-DV57-1
Enable Menu Item Description
This option controls the Auto-OFF feature, which automatically stops the fl ow of air
after taking off the mask. The device will require at least 30 seconds to turn OFF
depending on the kind of mask used. When Auto-OFF is enabled, Auto-ON is also
enabled. If the unlocked symbol is shown, you are able to adjust this setting. If the
locked symbol is shown, this setting can only be adjusted by your provider.
This option controls the Auto-ON feature, which automatically starts the fl ow of air
after breathing once or twice into the mask. This is always enabled when Auto-OFF is
enabled. If the unlocked symbol is shown, you are able to adjust this setting. If the
locked symbol is shown, this setting can only be adjusted by your provider.
This option controls the mask fi t check feature. The air fl ow to your mask is
constantly measured and if the amount of airfl ow exceeds 95 liters/minute for more
than 10% of the time used, a notifi cation will be created. This notifi cation will be
displayed the next time you use the SleepCube.
Low backlight Enable will keep the backlight dim during operation. Low backlight
Disabled will turn the backlight OFF during operation.
Pressing the up key exits the Enable menu.
EN - 9