Maintenance & Repair - Pari Vios LC PLUS Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

G. DISINFECTING YOUR PARI
REUSABLE NEBULIZER
1.
Clean the nebulizer as described in Section F . For
low-level disinfection every other treatment day, soak
all parts of the PARI Reusable Nebulizer (except
tubing and mask) in a solution of 1 part distilled
white vinegar and 3 parts hot tap water for 1 hour.
You can substitute respiratory equipment disinfectant
(like Control III
) for distilled white vinegar (follow
®
manufacturer's instructions for mixing).
2.
Rinse all parts of the PARI Reusable Nebulizer
thoroughly with warm tap water and dry with a clean,
lint-free cloth. Discard the vinegar solution when
disinfection is complete.
3.
The PARI Reusable Nebulizer parts (except tubing
and mask) may also be disinfected by boiling them in a
clean pot of water for a full 5 minutes.
4.
Hospitals and clinics can sterilize the PARI Reusable
Nebulizer parts (except mask and tubing) by
autoclaving (a maximum of 277° F or 136° C), follow
autoclave manufacturer's instructions.
5.
Dry parts on a clean, lint-free cloth.
CONTROL III
®
is a registered trademark of Maril Products, Inc.
NOTE: FOR PATIENTS WITH CYSTIC FIBROSIS, THE CF FOUNDATION
RECOMMENDS BOILING FOR A FULL 5 MINUTES TO DISINFECT.
H. MAINTENANCE OF YOUR VIOS
FILTER
Your Vios
air filter is bright white when new. With use,
®
the filter will slowly turn grey, indicating the need for
replacement. The filter must be replaced immediately if it
gets wet or becomes clogged. A good rule to follow is to
replace your filter every 6 months – at the same time as
your PARI reusable nebulizer.
Removing: First make sure the Vios
turned off. Using a coin, remove the air filter cover by
turning it counter-clockwise (Fig.E). Pull the cap and filter
out, then remove the filter from the cap (FigE-F).
F
®
power switch is
®
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOS
NOTE: THERE ARE TWO TYPES OF FILTERS THAT FIT
THE VIOS
COMPRESSOR. IF YOUR FILTER IS OPEN
®
ONLY ON ONE END AS SHOWN IN FIGURE F,
FOLLOW INSTRUCTIONS AS SHOWN IN FIGURE F
AND G. IF YOUR FILTER IS OPEN ON BOTH ENDS,
PRESS EITHER END OF THE FILTER INTO THE CAP.
Take a new filter and press into cap with the open end
visible. Place the cap and filter back into the front of
your Vios
. Use the coin to tighten by turning the cover
®
clockwise.
NOTE: DO NOT ATTEMPT TO WASH AND REUSE
AIR FILTERS. DO NOT USE SUBSTITUTE FILTER
MATERIAL COMPRESSOR DAMAGE MAY RESULT.
COMPRESSOR
Use a damp cloth to wipe the exterior of your Vios
housing and Wing Tip
tubing.
®
CAUTION
Never spray fluid into the ventilation slots on your Vios
Fluid could cause damage to the electrical & compressor
parts and could lead to malfunction. In the event fluid
enters the unit, please contact PARI Patient Connection at
1.800.FAST.NEB (327.8632).
MAINTENANCE & REPAIR
Never operate your Vios
®
incompletely assembled or damaged. Use only original
PARI parts and attachments authorized by PARI for the
Vios
Aerosol Delivery System. Use of other parts and
®
attachments may damage the unit or cause improper
treatment. Do not open or disassemble the compressor
unit at any time.
NOTE: THERE ARE NO SERVICEABLE PARTS INSIDE
THE VIOS
AEROSOL DELIVERY SYSTEM. SERVICE BY
®
ANYONE VOIDS WARRANTY.
G
®
if it is improperly or
EN-6
.
®

Publicidad

Tabla de contenido
loading

Tabla de contenido