Disinfection
Step 1
For disinfection, wipe all outside surfaces of the meter and lancing device thoroughly
with the bleach wipe until they are wet in order to disinfect the meter and lancing
device . Clean the strip port, data port and battery compartment with caution and do not
force liquids into sensitive meter electronic openings .
Step 2
Allow the meter and lancing device surfaces to remain wet for 60 seconds (1 full minute) .
After 1 full minute, immediately dry with drying towel to remove any residual moisture .
Disinfect your meter at least once a week . Disinfect your lancing device when visibly
dirty (at least once a week) .
Step 3
Allow the meter and lancing device to air dry completely before performing a blood
glucose test .
Step 4
When finished, thoroughly wash your hands with soap and water .