OPERATING INSTRUCTIONS
OVERLOAD
6.2 Overload indication. When the scale is overloaded, "OVER" will shows.
OVEr
6.3 Negative weight. When there is a negative weight value, "UNDER" shows and the scale will correct this
UndEr
error by rezero-setting.
7. POWER SOURCE
ZERO
7.1 6*AA batteries or a 9V adapter. When the adapter is used, the battery icon will not be displayed
0.0
C
KG
8. CALIBRATION
1. Press UP and Down at the same time, then press ON/OFF to start the scale. LCD will displays "CAL", then returns
to "0.0 LB"
2. Press DOWN key, it displays"-100", then put on 100lb weight. After display data becomes stable, Press DOWN key, it
displays "CAL" then "-200",
3. Put on 200lb weight. After display data becomes stable, press DOWN key, it displays "CAL" then "-300".
4. Put on 300lb weight. After display data becomes stable, press DOWN key,it displays "CAL" then "-400".
5. Put on 400lb weight. After display data becomes stable, press DOWN key, it displays "CAL" then "-500"
6. Put on 500lb weight. After display data becomes stable, press DOWN key, it displays "CAL" then auto off automatically.
ASSEMBLY INSTRUCTIONS
1. Connect Lower Pillar (B) to Base (C) using 4 Screws (D).
2. Connect Upper Pillar (A) to Lower Pillar (B).
3. Connect Upper Height Rod (E) to Lower Height Rod (F).
4. Place assembled height rod into slot (H) as shown. Pinch stops (G) to allow entry.
H
E
F
G
A
INTERNAL CABLE
CAUTION:
WATCH OUT FOR PINCH POINTS
DURING ASSEMBLY THAT MAY
DAMAGE INTERNAL CABLE!