The users of Unity are clinical specialists and end user. This device
should only be used with a prescription from a physician that includes
the necessary treatment information.
Note: The Unity System is designed to be used with Iceross Seal-In
liners, other applications are sole responsibility of the provider.
INDICATIONS
Prosthetic interface with suspension properties for a major lower limb
amputee; moderate impact use.
CONTRAINDICATIONS
• Amputee equipped with a loose fitting socket.
• The UNITY system must not be prescribed to those incapable of
utilising it according to instructions. For example those with poor
cognition.
WARNING!
Make sure that the socket is well fitting before any use; a loose fitting or
leaking socket may never be used with the Unity System.
STRUCTURAL SPECIFICATION
Net weight: approx. 75 g including tube.
Frame construction: Carbon fiber and aluminum frame.
Clearance and dimensions: See IFU for Pro-Flex LP.
FITTING THE UNITY SYSTEM
Note: To connect the system to a socket a Unity Socket Valve Kit is
recommended, the use of other applications is the sole responsibility of
the provider.
1. Fit the user with the foot containing the system
2. Cut the tube to correct length and connect to the Unity System and
the socket (using the Unity Socket Valve kit)
3. Use the supplied tube fixture to secure the tube against the pylon. The
tube should be located medially rather than laterally to avoid damage
during ambulation. Once the Unity System has been connected make
sure the pump is working properly; creating vacuum during gait and
not leaking air.
DISASSEMBLING / ASSEMBLING THE UNITY SYSTEM
The Unity System can be dissembled in case it needs to be cleaned or for
troubleshooting purposes.
Caution: The Unity System should only be disassembled / assembled by
a clinical specialist.
Caution: The inlet and outlet (Figure 4) ports on the membrane housing
should not be disassembled unless absolutely necessary. This part of the
system contains miniature duckbill valves that are easily lost and can
cause the Unity System to malfunction if not correctly reassembled.
HOW TO CHANGE MEMBRANE IN THE UNITY SYSTEM
1. Use the Flex-foot shoehorn (or equivalent) to remove the foot from
the foot cover.
2. Disconnect the tube from the Unity Membrane housing.
3. Untighten the screw that fastens the membrane housing to the Pro-
Flex LP pyramid.
4. Follow Figure 5.
®
6