17
Technical Information
Vent Accessories
Figure 37
Then screw the vent valve into the pump and tighten it using a 16 mm
hexagonal spanner with a torque of 2.5 Nm.
Figure 38
Do not overtighten the valve as this may damage the thread on the pump.
CAUTION!
Then connect the cable from the valve to the suitable connector on the
controller (see the preceding paragraph "INTERCONNECTIONS").
TV 1001 SEM User Manual / 87-900-973-01 (H)
277/296