2-7 Installation, MT36xx Series
The MT36xx Series is installed as follows:
a. Carefully remove the covers or plugs from the inlet and outlet of the
flowmeter.
b. The flowmeter must be installed with the inlet at the bottom and the
outlet perpendicular to the inlet. The indicator housing can be rotated
360°, for easy viewing, by loosening and then tightening the three set
screws at the base of the indicator housing.
c. When installing the flowmeter in the process line, follow accepted
plumbing practices.
d. It is strongly recommended that the bypass piping arrangement
shown in Figure 2-1 be used when installing the flowmeter. This
arrangement will allow the flowmeter to be removed for repair or
service without interrupting the flow stream.
e. Install the flowmeter within five degrees of true vertical. Use a level to
verify the alignment of the flowmeter. Accurate alignment to vertical is
required to maintain flowmeter accuracy.
HORIZONTAL
LINE
FLOWMETER
A
E
C
A - Inlet Valve
D - Control Valve
FIgure 2-1 Typical Installations
Section 2 Installation
VERTICAL
LINE
FLOWMETER
B
D
B - Outlet Valve
E - Drain Valve
MT36xx Series
D
B
C
A
E
C - Bypass Valve
2-3