5.2. All temperature chambers via the same Ethernet/RS422-RS485 converter
Controlling and
monitoring PC
™
e.g., with APT-COM
PC
Connection RJ45
Ethernet card
The optional extension cable 50m (5023-0117) can be connected behind the 5m cable (5023-
0095).
5.3. Bus termination / RS485 operation
For the connection of RS485 networks, DIP switches 1 and 2 must be set to ON
To connect temperature chambers from BINDER, the recommended operating mode of the
interface converter is termination enabled (BINDER factory setting).
Some controllers, like RD3, MB1, MB2, have an internal termination. When combining different
controllers, communication errors can occur which are indicated by APT-COM™ as "timeout" or
"CRC error". In this case it can be necessary to disable termination at the interface converter.
To disable termination, set DIP switches 6 and 7 in the interface converter to Off. You can safely
try out the correct setting. No damage of controller or interface converter will occur.
Connection of Ethernet/RS422-RS485 converter 09/2016
Ethernet
Ethernet/RS422-RS485
Converter
Ethernet
RS422/485
Cable 5 m Art. No. 5023-0095,
PATCH cable
Cable 15 m
Art. No. 5023-0036
optional: extension cable 50m
Art. No. 5023-0117