VERIFYING THE GAUGE CALIBRATION
2
Switch on the Elcometer AVU and
ensure that the appropriate dolly
adaptor is fitted and the measurement
units are the same as the Elcometer
510 (refer to the instructions supplied
with the Elcometer AVU).
3
Connect the Elcometer 510 actuator
(with skirt fitted) to the Elcometer AVU
dolly adaptor.
4
Press 'Ok' on the Elcometer 510 when
connected. The Elcometer 510
automatically starts to apply pressure
until the first test load is reached.
5
Compare the test load with the reading
on the Elcometer AVU display. If the
Elcometer AVU reading is within the
acceptable range, displayed in brackets
underneath the test load, press 'Ok' to
proceed to the next test pressure and
repeat step 4. (If outside the acceptable
range, re-calibration is recommended.
Press 'Escape' to exit the calibration
verification procedure and contact
Elcometer or your local supplier for
further information).
6
When the final test load has been
reached
, if it
range, press 'Verify' to update the gauge
or 'Escape' to cancel. The date and time
of the last verification procedure is
recorded against each batch and can be
viewed via Batch/Review Batch/Batch
Information.
Note: The acceptable range is based on the 'system'
accuracy - the accuracy of the Elcometer 510 and
Elcometer AVU unit combined. Measurement
verification points: 20mm Dolly; 5, 15 & 25MPa,
50mm Dolly; 0.8, 2.4 & 4.0MPa
(or equivalent units).