"Service completed" displays the amount of service work performed
using the air conditioning service unit since the last reset (see date).
9.4.1
Entering the filter code
To reset the filter counter, you need to enter a 12-digit code. This special
code is located on the new filter. A "Code entry" prompt appears in the
menu. If a code is not entered, the unit is taken out of operation. Each
code only works once.
9.4.2
Leak test
After changing the filter, perform a leak test with an approved testing
device.
9.5

Calibrating the pressure transducer

I
NOTE
The pressure transducer has to be correctly calibrated in order to make
accurate pressure measurements.
Calibration is required:
 Every four weeks
 If the AirConServiceCenter has been subjected to heavy shaking
 Every time the vacuum pump oil is changed
 If the display shows implausible pressure readings
1. In the basic menu, use the cursor keys  or  to select "Other
selections":
2. Depressurise the unit (as described in chapter "Changing the dryer
filter" on page 110).
3. Press ENTER to confirm.
4. Use the cursor keys  or  to select "Service":
EN
Service completed:
Number
STOP-EXIT
Short selection
Free selection
Other selections
ENTER-OK
Int. vessel filling
Flushing
Reset scales.
Service.
Service tasks
08/03/17
4
113