a Use either a pre-moistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
from the list of approved cleaners.
Approved cleaners/disinfectants for the ultrasound system
Cleaner/disinfectant
SaniCloth AF3 (gray top)
SaniCloth Plus (red top)
PI-Spray II
1
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum
period of time.
2
Qualified for use as an intermediate-level disinfectant for mycobacteria. For a more complete list, refer to the
cleaners and disinfection document available at www.sonosite.com.
b Remove all gel, debris, and bodily fluids from the system.
c With a new wipe, clean the system, including the display, by wiping from clean areas to the soiled areas.
This method helps to avoid cross-contamination.
d Observe the minimum wet contact time recommended by the manufacturer. Monitor the system for
wet appearance. Re-apply with a new wipe if no longer wet.
e Allow the ultrasound system to air dry in a clean, well-ventilated space.
6 Clean the ultrasound system STAND to remove any debris or bodily fluids. For information about cleaning
the stand, see the Edge Stand and H-Universal Stand User Guide.
Cleaning and disinfecting
Compatible systems
2
Sonosite S Series
SonoSite Edge
SonoSite M-Turbo
SonoSite S Series
Minimum wet contact time
3 minutes
3 minutes
10 minutes
1
11